Table 1 |
|||||||
|
Testing results for Coartem/Riamet by expiry date for Raman spectrometry, TLC and disintegration |
|||||||
| July 2008 Test Resultsa |
February 2009 Test Resultsa |
||||||
|
|
|||||||
| Expiry Date |
Number of Samples |
Number of Months Past Expiry Date |
TLC |
Disintegration |
Raman |
Number of Months Past Expiry Date |
Raman |
|
|
|||||||
| June 2008 |
7 |
1 |
7/7 |
7/7 |
7/7 |
8 |
7/7 |
|
|
|||||||
| May 2008 |
39 |
2 |
39/39 |
39/39 |
39/39 |
9 |
39/39 |
|
|
|||||||
| February 2008 |
4 |
5 |
4/4 |
4/4 |
4/4 |
12 |
4/4 |
|
|
|||||||
| August 2007 |
7 |
11 |
7/7 |
7/7 |
7/7 |
18 |
7/7 |
|
|
|||||||
| May 2007 |
2 |
14 |
2/2 |
2/2 |
1/2 |
21 |
1/2 |
|
|
|||||||
| November 2006 |
3 |
20 |
3/3 |
3/3 |
3/3 |
27 |
3/3 |
|
|
|||||||
| October 2005 |
5 |
33 |
5/5 |
5/5 |
5/5 |
40 |
5/5 |
|
|
|||||||
| September 2003 |
1 |
58 |
1/1 |
0/1 |
1/1 |
65 |
0/1 |
|
|
|||||||
| Total |
68/68 |
67/68 |
67/68 |
66/68 |
|||
|
|
|||||||
|
a. Results are total number of samples that passed testing/total samples tested. |
|||||||
|
Bate et al. Malaria Journal 2009 8:33 doi:10.1186/1475-2875-8-33 |
|||||||