Evaluation of the SD FK70 Malaria Ag Plasmodium vivax rapid diagnostic test in a non-endemic setting

Philippe Gillet1, Katrien Bosselaers1, Lieselotte Cnops1, Emmanuel Bottieau1, Marjan Van Esbroeck1 and Jan Jacobs1,2*

Author Affiliations

1 Department of Clinical Sciences, Institute of Tropical Medicine (ITM), Antwerp, Belgium

2 Department of Medical Microbiology, Faculty of Health, Medicine and Life Sciences (FHML), Maastricht, the Netherlands

For all author emails, please log on.

Malaria Journal 2009, 8:129 doi:10.1186/1475-2875-8-129

Published: 11 June 2009

Abstract

Background

For clinical and epidemiological reasons, it is interesting to diagnose non-falciparum malaria to the species level. This retrospective study assessed the performance of the SD BIOLINE Malaria Antigen Pv test (FK70), a two-band immunochromatographic test detecting Plasmodium vivax-specific lactate dehydrogenase, on samples of international travellers in a non-endemic setting.

Methods

Stored blood samples from international travellers suspected of malaria were used, with microscopy corrected by PCR as the reference method. Samples infected by Plasmodium vivax (n = 100), Plasmodium falciparum (n = 75), Plasmodium ovale (n = 75) and Plasmodium malariae (n = 25) were included, as well as 100 malaria-negative samples. End points were sensitivity, specificity, inter-reader reliability and reproducibility.

Results

The overall sensitivity of the FK70 for the diagnosis of P. vivax was 88.0% (95% confidence interval (CI): 83.6% – 90.3%). For parasite densities > 500/μl, a sensitivity of 97.2% (CI: 92.6% – 99.1%) was obtained. Specificity was 98.5%, with 4 out of 75 P. falciparum samples testing positive. None of the P. ovale samples tested positive. Nearly two-thirds (57/88, 64.7%) of positive P. vivax samples showed faint or weak line intensities, with stronger line intensities at higher parasite densities. The test showed excellent reproducibility and reliability for test results and line intensities (kappa values exceeding 0.98 and 0.87 respectively).

Conclusion

The FK70 test performed well in diagnosing P. vivax infections in a non-endemic reference setting. It can be of added value to microscopy in species differentiation of malaria infections, especially at parasite densities > 500/μl.